When the Device is turned on and in RATE or MAX mode, the current
radiation reading and other log data is recorded at the set interval.
In SCALER Mode:
o
Auto log will be paused and will only log at the completion of each
count.
o
For a continuous Scaler count (Scaler time is 0), logs will be
recorded at auto log interval.
The Data Log Table is linear; once the table is full, no further writes will be
allowed until the Data Log Table is erased.
Note:
A short press of the LOG button will start a count and
automatically log the result. An extra-long press of the LOG
button during the count will toggle the display between the timer
and reading.
Format
The datalogging format is as follows:
Format Version (1 Byte)
Month (1 Byte)
Year (2 Bytes)
Day (1 Byte)
Hour (1 Byte)
Minutes (1 Byte)
Seconds (1 Byte)
Logged Value (4 Bytes)
Range (1 Byte)
Units (1 Byte)
Mode (1 Byte)
Detector Number (1 Byte)
Status (1 Byte)
Reserved (2 Bytes)
Elapsed Count Time in seconds (4 Bytes)
Scaler Count Time in seconds (4 Bytes)
Location ID (32 Bytes)