3-10
P/N 9-500-0255
Service Manual
Chapter 3: Functional Checks/Setting Defaults
The Affinity Series Control Panel
Exposure Default Screen (Figure 3-9)
allows the user to set system defaults
for exposure techniques. The choices
made in the Exposure Default Screen
will be reflected in the Run Mode
Exposure settings displayed after
power up.
Press
Reset
+
Right Arrow
together to select the User Default Screen from the Run Mode
Screen. To select the Exposure Default Screen from the Run Mode Screen, press
Reset
+
Right Arrow
twice. To go back to the Run Mode Screen from the Exposure Default Screen,
press
Reset
+
Left Arrow
twice, from the User Default Screen, press
Reset
+
Left
Arrow
once.
Note…
Access to the Calibration Mode (the
Debug Screen and the three
Calibration Screens) will require the
servicing engineer to configure a
DIP switch on the Microprocessor
Board (see Chapter 4, Section 2.1,
Entering Calibration Mode).
The Calibration Mode/Debug Screen
(Figure 3-10) displays serial communi-
cations information throughout the unit.
Figure 3-9: The Exposure Default
Screen
Figure 3-10: The Calibration Mode/
Debug Screen
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...