5-10
P/N 9-500-0255
Service Manual
Chapter 5: Maintenance—General
Troubleshooting
3.3
System Data Retrieval
The Affinity System provides the Service Engineer with the capability to download/upload
calibration data and the Alert Codes Log to a laptop computer with a Terminal Emulation
Program (such as HyperTerm). This allows the Service Engineer to print or append this data
as needed.
In order download this information, a Null-Modem, 9 pin to 25 pin Adapter, approximately
6 feet, is required (Tandy Radio Shack Catalog #950-0188, or equivalent).
1. Connect the laptop computer to the Affinity System using the Null-Modem Cable as
shown in Figure 5-1.
85
BOOST_TUBEHEAD_COMM_ERROR
SYSTEM ERROR
During boost, Host board is unable to
communicate with Tubehead Controller.
86
AF_EXPOSURE_TUBEHEAD_COMM_ERRO
R
SYSTEM ERROR
During Auto-Filter Exposure, Host board
is unable to communicate with Tube-
head Controller.
90
FLASHER_COMM_ERROR
AUTO ID
Flasher does not respond.
92
PRINTER_NO_PAPER_ERROR
PRINT NO PAPER
93
PRINTER_OFF_LINE_ERROR
PRINT OFF LINE
95
INTERLOCK_STATUS_ERROR
INTERLOCKS
Interlock circuit hardware reports an
error.
Table 5-2: Affinity Series Alert Messages
No.
Alert Name
Alert Messages
Comment
Table 5-3: Null-Modem, 9-pin to 25-pin Adapter Connections
9-pin, Female, DB-9
25-pin, Male, DB-25
Signal
From
To
Signal
RX Data
2
2
TX Data
TX Data
3
3
RX Data
S-Ground
5
7
S-Ground
CTS
7
5
CTS
RTS
8
4
RTS
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...