![LORAD AFFINITY Series Скачать руководство пользователя страница 139](http://html.mh-extra.com/html/lorad/affinity-series/affinity-series_service-manual_1939281139.webp)
P/N 9-500-0255
6-11
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
Gantry Components—Remove and Replace
Affinity Serial Numbers 0001 through 0018
1. Remove the HV Generator Assembly as follows (refer to Figure 6-7):
A. Remove power to unit and access the High Voltage Generator Assembly by
removing the left side Gantry cover as per Section 2.1.1.
B. On the top of the HV Multiplier, disconnect the HV cable from the HV
cable receptacle.
C. On the HV Inverter board, disconnect GJ1. Wait approximately 6 minutes
before proceeding (see WARNING below).
WARNING!
To reduce the risk of hazardous electrical shock, do not
attempt service until the LEDs D2 (Inverter Rail) and D35
(Brake Rail) on the HV Inverter Board are extinguished (this
will take approximately 5 minutes). Once the LEDs are
extinguished, do not proceed until an additional 1 minute has
elapsed.
D. On the HV Inverter board, disconnect GJ2, GJ3, and GJ5. Remove any
installed cables ties.
E.
Remove the Ribbon Cable between the HV Inverter board and the HV Con-
trol Filament board. Set aside for re-installation later.
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...