P/N 9-500-0255
A-7
Service Manual
Appendix A: Specifications
System Specifications
Over-Temperature Protection Sensor
Internally provided in series with Stator com-
mon. NC 10 Amp @240 VAC, open @ 79.4° C
(174.9° F), close @ 68.3° C (154.9° F)
Safety Class
IEC 601-2-28
X-ray Tube Head Cooling
A fan is provided to maintain a tube housing
temperature of no greater than 85° C (185° F)
with an ambient temperature of 25° C (77° F).
A fan is provided to keep the lamp housing at
a temperature no greater than 65° C (149° F).
Added Filtration
The system provides a 30 micron (0.03 mm)
Molybdenum foil filter and a 25 micron (0.025
mm) Rhodium foil filter. The filter movement is
automatic and has position sensors for use as
x-ray interlock.
Beam Quality
With the Compression Paddle in the x-ray
beam, the HVL for Mo/Mo operation is equal
to or greater than kV/100 +0.03 mm Al but
less than kV/100 +0.12 mm Al. For Mo/Rh,
minimum value of > kV/100 +0.03, Max < kV/
100 +0.19 mm Al.
X-ray Collimation
The system detects the attached image recep-
tor and full field collimation is provided by an
auto-aperture mechanism with additional
fixed apertures for magnification and reduced
field operation. An aperture detection system
ensures that the correct aperture is used with
the installed image receptor. Exposures are
inhibited when improper combinations are
detected.
The collimator has a beam-limiting device that
allows the entire chest wall edge of the x-ray
field to extend 6.5 cm, +2.5 cm, beyond the
edge of the Image Receptor and provides a
means to ensure the x-ray field does not
extend beyond any chest wall edge of the
Image Receptor of more than 2% of the SID.
Light Field to X-ray Field Congruency
The light field that passes through the x-ray
beam limitation device is aligned with the x-
ray field so that the total of any misalignment
of the edges of the light field and the x-ray field
along either the length or the width of the visu-
ally defined field at the plain of the breast sup-
port surface does not exceed 2% of the SID.
Light Field Lamp
The lamp illuminates for 30 seconds ±5 sec-
onds upon pressing a Light Field switch on the
C-arm, or by pressing a compression Down
switch. The lamp extinguishes automatically
when x-ray exposure is initiated. A shatter
shield is provided.
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...