![LORAD AFFINITY Series Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/lorad/affinity-series/affinity-series_service-manual_1939281137.webp)
P/N 9-500-0255
6-9
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
Gantry Components—Remove and Replace
2.3.1
Display Microprocessor Board
Perform this procedure to remove and replace the Display Microprocessor Board
(Figure 6-6, Item 1). Instructions are also provided to remove the firmware chip (U2)
which contains the embedded user interface software.
1. Remove power to the unit. Disconnect the Control Panel Assembly from the
mount on the side of the Gantry, or from the remote location mount.
2. Turn the Control Panel upside down. Remove the 4 screws that secure the rear
cover to the front cover. Lift the rear cover off of the assembly.
3. Disconnect the input (data and power) wire harnesses from their jacks at BJ1
and BJ8. Pull the rear cover off of the wire harness and store it in a safe location.
4. If replacing the entire Microprocessor Board, skip to Step 6. If replacing just the
firmware chip, proceed to Step 5.
5. Locate the firmware chip (U2) on the Microprocessor Board. Use a PLCC
extraction tool to remove the chip. Install the replacement chip into the socket.
Be sure to line up the dot on the chip with the arrow in the socket. Skip to Step
9.
6. Disconnect the keypad ribbon cable from the jack at BJ3. Disconnect the
display back light cable from BJ6.
7. Remove the four screws that secure the Microprocessor Board to the front cover.
Lift the board off the cover, then disconnect the display data cable from BJ2 on
the opposite side of the board.
8. Install the replacement Display Microprocessor by reversing this procedure.
9. When complete, install the rear cover, then remount the Control Panel to the
unit, or to the remote mounting location.
10. Power up the unit and verify that all the screens are available. Also verify that all
of the keypad functions are operational.
2.3.2
LCD Display
Perform this procedure to remove and replace the LCD Display (Figure 6-6, Item 2).
This procedure requires that the Display Microprocessor be removed.
1. Remove power to the unit.
2. Disconnect the Control Panel Assembly from the mount on the side of the
Gantry, or from the remote location mount.
3. Remove the Display Microprocessor Board per Section 2.3.1.
4. Remove the 4 screws that secure the LCD Display to the front cover.
5. Install the replacement LCD Display by reversing this procedure.
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...