P/N 9-500-0255
6-51
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
C-Arm Components—Remove and Replace
6. Install a compression paddle and check that system status changes from
“Standby” to “Ready”. Remove the paddle and verify that system status returns
to “Standby”.
3.3.6
Compression Thickness Potentiometer
The Compression Thickness Potentiometer is a 5,000-ohm, 10-turn potentiometer
that mounts to the left side bulkhead of the compression device frame.
1. Turn power OFF.
2. Remove any installed compression accessory.
3. Remove the C-arm bottom cover and the left side cover as per Section 3.1.2.
4. Remove the Tubehead covers as per Section 3.2.1.
5. Remove the top and bottom compression device covers (Section 3.3.1). Use
care not to damage the compression device bellows.
6. Loosen the two allen screws securing the Compression Timing Belt to the
Tubehead Support Plate (Figure 6-35, Item 1). Remove timing belt from the
upper clamp block (Figure 6-35, item 2).