P/N 9-500-0255
B-7
Service Manual
Appendix B: Technical References
LED Indicators
Table B-14: C-arm Safety Board (PN 1-003-0440) LED Indicators
Table B-15: Relay Protection Board (PN 1-003-0490) LED Indicators
Table B-16: Host Microprocessor Board (PN 1-003-0493) LED Indicators
Table B-17: Host Microprocessor Board—C-arm Switch Multiple Function Indicators
LED
LED “ON” Indication
LED
LED “ON” Indication
D1
+5 V
D6
ACC0
D2
Actuator Fault
D7
ACC1
D3
Load Cell Fault (load cell at limit)
D8
ACC2
D4
Clock signal (flickers to indicate board is working) D9
VCC
D5
CAL Mode Initiated
D10
-5 V
LED
LED “ON” Indication
D1
No power to Board
LED
LED “ON” Indication
D1
+5 V
D2
+12 V
D3
-12 V
D4
LED Serial CLK (Flickers when operating)
D5
C Angle (Flickers when operating)
D6
MOSI (Flickers when operating)
D7
COMP MISO (Flickers when operating)
D8
COMP SS (Flickers when operating)
D9
TUBE SS (Flickers when operating)
D15
Unit set to Large Focal Spot
D16
Unit in Boost mode
D17
Unit in BRAKE mode
D18
Unit in RUN mode
D19
Filament Enabled
D25
Rotation Brake Released
D26
Actuator Down
D27
Actuator Up
C-Arm Switch Function
D21
D22
D23
D24
Compression Release Active when:
ON
ON
OFF
OFF
Compression Down Active when:
OFF
ON
OFF
ON
Compression Up Active when
ON
ON
OFF
OFF
Compression Brake Engaged when
OFF
OFF
ON
ON
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...