Measuring
35
3 variants are available:
1 = Delivery state (127/220/380 V)
2 =
New mains voltage Europe (110/230/400 V)
3 = Voltage actually present during measurement
Modification of the instrument to other mains voltages can be done in a service
facility.
If the actual mains voltage is used (variant 3) for the display of the short-circuit
current this display is correct for the momentary state but has not been determined
in conformity with standards. In this case, variant 2 should be chosen.
Note:
For highly accurate measurements of low loop and internal network resistances
below 1 Ohm, the measuring-lead zero compensation may be performed as per
section 3.10.
Error Messages:
Flashing symbols on the display generally signify illegal conditions or errors, for
further information se section 3.11 „Display hints“.
Содержание Saturn 100 Plus
Страница 2: ......
Страница 7: ...Initial power up 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Страница 13: ...Measuring 11 Flowchart of operation...
Страница 46: ...Information for the tester 44 4 2 Description of Function...
Страница 71: ...Erste Inbetriebnahme 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Страница 76: ...Messfunktionen 10 Flu diagramm f r Bedienung...
Страница 107: ...Informationen zum Pr fger t 41 4 2 Funktionsbeschreibung...
Страница 126: ...Service 60...
Страница 131: ...Premi re mise en service 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Страница 138: ...Directives d utilisation 12 Organigramme d utilisation...
Страница 174: ...Informations 48 4 2 Les fonctions...
Страница 199: ...Primo Avvio 6 9 2 1 10 5 4 6 7 8 3 11 A 12...
Страница 204: ...Funzioni di misura 11 Diagramma di flusso per l utilizzo...
Страница 263: ...Puesta en marcha 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Страница 269: ...Medici n 11...