9. Contrast Methods
9.1.1 Brightfield
• If present: click the condenser disk* into the
BF position.
• If present: pull the light ring slide* out of the
condenser.
• If present: switch the fluorescence illuminator
into an empty position or filter system A.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
• Movable condenser heads:
The condenser top is swung out for
objective magnifications < 10x.
• Bring the image into focus using the focus dial
and set the brightness.
• For an optimal field diaphragm setting, check
the Köhler illumination (
→
p. 31).
• Use suitable transmitted-light filters as
applicable (Fig. 47).