8. Operation
Attention:
The aperture diaphragm in the
illumination light
path
is
not
for setting the image brightness. Only
the rotary brightness adjustment knob or the
neutral density filter should be used for this.
An aperture diaphragm in the
objective
is
normally fully opened. The reduction in image
brightness caused by stopping down results in:
Greater depth of field
Less coverglass sensitivity
Suitability for darkfield
Change in contrast
8.9 Field diaphragm
The field diaphragm (46.5) protects the
specimen from unnecessary warming and keeps
all light not required for image formation away
from the object to enable greater contrast. It is
therefore only opened just wide enough to
illuminate the viewed or photographed object
field. A change in magnification therefore
always necessitates matching of the field
diaphragm.
Color-coded Condenser
The color markings on the condenser (46.2)
correspond to the color rings of the objectives.
When changing objectives, a suitable aperture
diaphragm setting can be found by setting it to
the matching color marking (corresponds to 2/3
of the objective-side aperture).
Fig. 46
CL/PH condenser
1
Slot for light rings, etc.
2
Color coding
3
Aperture diaphragm
4
Filter holder
5
Field diaphragm
2
3
1
4
5
Содержание DM1000 LED
Страница 1: ...Leica DM1000 Operating manual Bedienungsanleitung Mode d emploi M I C R O S Y S T E M S...
Страница 3: ...3 Leica DM1000 Operating Manual...
Страница 61: ...3 Leica DM1000 Bedienungsanleitung M I C R O S Y S T E M S...
Страница 119: ...3 Leica DM1000 Mode d emploi...
Страница 177: ......