19
6. Assembly
6.3 Tube and Eyepieces
Note:
For fluorescence applications, install the
fluorescence illuminator first
→
p. 21.
If available, the analyzer* (10.1) must be inserted
into the stand. This requires that the guide key
engages in the guide pin (10.2).
To mount the analyzer, the analyzer mount TL*
20 mm or 60 mm can also be placed between
stand and tube.
An intermediate tube pole* with a switchable
analyzer (on/off) and Bertrand lens is also
available as an option.
The tube is mounted to the stand either directly or
with the use of intermediate modules.
Fig. 11
Fastening the tube
1
Clamping screw
1
!
Fig. 10
Analyzer mounting
1
Analyzer
2
Orientation pin and guiding notch
3
Clamping screw
• Loosen the clamping screw (11.1) on the
stand.
• Insert the tube in the circular receptacle
(dovetail ring).
• Retighten the clamping screw (11.1).
• The eyepieces are inserted into the eyepiece
tubes on the tube.
6.4 Objectives
Always only use Leica objectives of tube length
×
(infinity)! The standard thread is M25. The
objectives should be arranged so that the
magnification increases when the objective
nosepiece is rotated counter clockwise.
Attention:
Lower the specimen stage as far as possible
before assembling the objectives. Close vacant
threads in the nosepiece with dust protection
caps!