6. Assembly
6.5 Light Source for the Transmitted Light Axis
Caution!
Be sure that the lamp housing is discon-
nected from the power supply. Unplug the
power plug and the power supply during as-
sembly.
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Replacing the Lamp of the Integrated
Illumination
The transmitted light illumination with a low-
voltage tungsten halogen lamp (Fig. 12) is
integrated in the base of the microscope and is
accessible from the right-hand side.
• Remove the insert (12.2).
Caution!
The lamp may still be hot!
• Remove the lamp.
Caution!
Do not remove the new lamp’s dust cover
until you have installed the lamp. Avoid
fingerprints on the lamp.
• Insert the new lamp with the dust cover
straight into the socket until it stops. Be sure
that the lamp is inserted straight.
• Remove the lamp’s dust cover.
• Replace the insert (12.2).
Fig. 12
Transmitted-light illumination in microscope base
1
Tungsten halogen lamp
2
Insert
1
2