Troubleshooting
7
7.3
Malfunctions of the two-in-one blade holder E
7.3.1 Replacing the pressure plate
Fig. 81
Fig. 82
1
2
3
2
3
4
5
1. Bring the safety guard down
.
2. Rotate the clamping lever counterclockwise
.
4. Remove the damaged pressure plate
.
Tips
Make sure that the insert of the clamping top is installed as shown
in case it
falls out. If it is installed wrongly, the pressure plate cannot be clamped.
5. Install a new pressure plate.
6. Insert the bolt
7. Rotate clamping lever clockwise
Содержание HistoCore BIOCUT