Version 1.7, Revision K
Operation
5
Warning
The rotation speed of the handwheel does not suit the hardness of the specimen.
Instrument damage and possibly specimen damage.
• The rotation speed of the handwheel must be adapted to suit the hardness of the specimen. For
harder specimens, use a slower speed.
Warning
In the manual sectioning mode, the user cuts the specimen block by rotating the handwheel at a very
high speed.
It may lead to poor sectioning quality and even damage the specimen.
• In the manual sectioning mode, the rotation speed of the handwheel shall be no more than 60 rpm.
Caution
After the knife/blade holder is laterally displaced, the specimen is not retracted and retrimmed.
Specimen damage.
• Retract the object head and retrim the specimen block each time after laterally displacing the
knife/blade holder.
Note
Accessories and components experience corrosion due to using corrosive/strongly acidic/alkaline
reagent or solvent with the instrument or accessories, such as decalcified solution that contains
acid, ammonium hydroxide that contains alkali, etc.
Accessories can malfunction.
• Avoid corrosive/strongly acidic/alkaline reagent or solvent dripping onto the instrument surface or
accessories.
• If such reagent or solvent has dripped onto the instrument surface or accessories, wipe off the
residue and dry the accessories sufficiently as soon as possible.
• If such reagent or solvent is used frequently, carry out a thorough daily cleaning of the blade
holder, the universal cassette clamp (UCC), and other accessories if necessary.
1. Set the required section thickness, or check the value setting on the window for displaying the
section thickness, respectively.
2. Always use a different area of the cutting edge for trimming and sectioning. To do so, laterally
displace the knife in the knife holder. When using the two-in-one blade holder E with lateral
displacement, the complete clamping top is moved.
3. For sectioning, turn the handwheel evenly in a clockwise direction.
Caution
Rotating the handwheel and the coarse feed wheel at the same time.
Specimen damage.
• Do not turn the coarse feed wheel whilst rotating the handwheel.
4. Pick up the sections, stretch them and mount them on microscope slides.