39
HistoCore BIOCUT
Operation
5
5.6
Clamping the knife/disposable blade
Warning
Knives or blades are handled and/or disposed in an inappropriate manner.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Work especially thoroughly and carefully when handling knives and/or blades.
• Always wear dedicated clothes (including cut-resistant gloves) when handling knives and/or
blades.
• Dispose knives and blades always in a safe place (for example, dedicated knife case) and an
appropriate manner that ensures that people cannot get hurt.
• Never place a knife anywhere with the cutting edge facing upwards and never try to catch a
falling knife!
• Always clamp the specimen BEFORE mounting and clamping the knife/blade.
Warning
The operator installs two knives and/or blades on the knife and/or blade holder.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Do not clamp two knives/blades on the knife holder/blade holder. Install the knife/blade in the
center of the knife/blade holder. Knives/Blades must not exceed the edge of the knife/blade
holder.
Warning
Too thick or thin blade used in the two-in-one blade holder.
Sample impaired.
• Use a blade with a maximum thickness of less than 0.322mm and a minimum thickness of more
than 0.246mm.
• The recommended compatible blade is listed in Optional Accessories
.
Warning
The blade is not installed completely parallel to the upper edge of the pressure plate due to excess
paraffin, uncleaned inserts, etc..
When using the lateral movement function, failure to install the blade parallel to the pressure plate
may result in poor sectioning results. For example, if a section is too thick or thin, chattering within
the section, in the worst-case scenario, may damage the specimen.
• Do not continue the sectioning when getting unsatisfying section results.
• Re-install the blade and make sure that it is parallel to the upper edge of the pressure plate.
• Always check the parallelism between the blade and pressure plate after moving the blade using
the lateral movement function.
• Ensure there is no paraffin waste and the insert is clean before use.