56
Version 1.7, Revision K
Optional Accessories
6
6.3.2 Knife holder base, without lateral displacement feature
Fig. 39
1
2
Repositioning the knife holder base
The one-piece knife holder base (rigid)
can be moved forwards and backwards on
the microtome base plate.
This vertical displacement allows bringing the
knife holder into the optimal cutting position in
relation to the specimen.
• To release, rotate the clamping lever
on the right side of the microtome base
plate counterclockwise.
• Reposition the knife holder together with the
knife holder base forward or backward as
appropriate.
• Secure the clamping mechanism by rotating the
lever
6.3.3 Knife holder E with a water trough for low-profile blades
Fig. 40
3
1
2
The knife holder E with a water trough
is only for low-profile blades.
The safety guard on knife holder E consists of a
red foldaway handle
. To cover the
cutting edge, fold the safety guard handle
clockwise as illustrated in the figure.
Tips
The clamping levers on the knife holder are not interchangeable. The two clamping levers (
) must remain in the position shown at all times, as otherwise isolated
malfunctions of the knife holder can occur.
Clamping lever for the blade
is at the right, clamping lever for the lateral displacement
is at the left.