14
Version 1.7, Revision K
Safety
2
Warning
The operator grabs into the knife/blade due to inappropriate working procedure.
Serious injury of persons when putting the specimen in place if knife/blade has been installed
before.
• Prior to uploading the specimen on the microtome, make sure that the cutting edge has been
covered by the safety guard and the hand wheel locking mechanism has been activated. Once
the operator wants to upload the specimen as well as inserting a knife/blade, always upload the
specimen block before mounting and clamping the knife/blade.
Warning
The operator installs two knives/blades on the knife/blade holder.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Do not clamp two knives/blades on the knife holder/blade holder. Install the knife/blade in the
center of the knife/blade holder. Knives/blades must not exceed the edge of the knife/blade
holder.
Warning
Too thick or thin blade used in the two-in-one blade holder.
Sample impaired.
• Use a blade with a maximum thickness of less than 0.322mm and a minimum thickness of more
than 0.246mm.
• The recommended compatible blade is listed in Optional Accessories
Warning
The blade is not installed completely parallel to the upper edge of the pressure plate due to excess
paraffin, uncleaned inserts, etc..
When using the lateral movement function, failure to install the blade parallel to the pressure plate
may result in poor sectioning results. For example, if a section is too thick or thin, chattering within
the section, in the worst-case scenario, may damage the specimen.
• Do not continue the sectioning when getting unsatisfying section results.
• Re-install the blade and make sure that it is parallel to the upper edge of the pressure plate.
• Always check the parallelism between the blade and pressure plate after moving the blade using
the lateral movement function.
• Ensure there is no paraffin waste and the insert is clean before use.
Warning
The knife/blade is installed before the knife/blade holder and the knife/blade holder base are installed
on the instrument.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Prior to inserting the knife/blade, both the knife/blade holder and the knife/blade holder base must
be installed on the microtome.