Repeat this Section until no further adjustment is necessary.
Press BACK button on the test key pad to get back to the main (frequency) screen.
tone=0 (turns Off test tone)
Low pass filter set / and Frequency adjust
Prerequisite(s) Unit in set-up mode
Unit at middle carrier frequency
Note
Entering Set-up mode is not necessary if the Alternate Method is exercised.
Using the test panel key pad navigate to the "Rolloff" screen LCD reads "Rolloff 35Hz" If LCD reads anything other
than "Rolloff 35Hz" press the DWN button on the test panel key pad to change it to "Rolloff 35Hz".
rolloff=0 (set LF roll-off to 35 Hz)
Using the test panel key pad navigate to the "Fine" screen (menu function that allows offset of carrier frequency)
Press UP or DWN button on the test panel key pad to fine adjust middle carrier
frequency
+/- 2kHz
finetune? (gets setting) finetune=xx (set fine frequency correction -16 to
+15, 3.125 kHz steps)
Pilot signal deviation & frequency measurements
Prerequisite(s) no audio in at audio input rig
Gain set to "0"
level=0 (changes audio level to 0)
Using the test panel key pad navigate to the "freq." screen
Press MENU/SEL and UP button simultaneously to change frequency to the lowest carrier frequency. (pressing
MUNU/SEL and UP simultaneously will step the frequency up in increments of 16 only pressing the UP or DOWN
buttons will only increment by 1) .
channel=XX
set channel (0(00) to 1020 (ff) in normal
tuning mode, 80 is 512)
Measure peak carrier deviation at radio board J3
4.5 to 6.5 KHz
(5.5)
2.8 to 3.4KHz
(3.1)