![LeCroy HDO6000 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/lecroy/hdo6000/hdo6000_operators-manual_1866315047.webp)
Operator's Manual
V
IEW
S
EGMENT
A
S
M
ATH
F
UNCTION
Besides using the Zoom feature, you can also create a Math (F
x
) trace to display individual segments.
1. From the menu bar, choose
Math
→
Math Setup...
2. Touch a
Function (F
x
) tab
to display its corresponding dialog.
3. On the dialog, touch
Operator1
and select the
Segment
button from the pop-up menu.
4. Touch the
Select
right-hand dialog tab.
5. Touch
First Selected
and choose the first segment to display.
6. Touch
Number of Selected
and enter the number of segments to display at once.
V
IEW
S
EGMENT
T
IME
S
TAMPS
To view time stamps for each segment:
1. From the menu bar, choose
Timebase
→
Acquisition Status
or
Vertical
→
Channel Status
.
2. Touch the
Trigger Time
tab.
3. Under
Show Status For
, choose
Time
.
4. In
Select Segment
, enter the segment number of interest.
You can also touch the Up/Down Arrow buttons to scroll through segment times.
Clock Source Settings
Sample Clock
These settings determine the clock that controls when the oscilloscope's digitizers sample the input
waveforms. The default setting is to use the oscilloscope's
Internal
clock. To use an external sample
clock:
1. Connect the clock source to the
Ext
input on the front of the oscilloscope using a BNC cable.
2. Go to
TImebase > Horizontal Setup
and choose
Real-time Sampling Mode
.
3. On the
Clock Source tab
under
Sample Clock
choose from 0V, ECL, TTL.
4. Choose an
External Coupling
of 50 Ohms, Ground, or 1 M Ohm.
Reference Clock
These settings control the Timebase reference used to synchronize acquisition across all channels. The
default setting is to use the oscilloscope's
Internal
10 MHz clock. To use an external reference clock:
1. Connect the clock source to the
Ref In/Out 10 MHz
input on the back of the oscilloscope using a BNC
cable.
2. Go to
Timebase > Horizontal Setup
and choose
Real-Time Sampling Mode
.
3. On the
Clock Source tab
under Reference Clock choose
External
.
922499 Rev B