HDO6000 High Definition Oscilloscope
NOTE
: Recovery takes approximately 4 to 15 minutes depending on the version of operating system
that is being restored. The progress is displayed on the screen.
11. When recovery is complete, you will see the message "Recover operation succeeded." Click
OK
.
12. Click the
close button
to exit the Acronis window. The oscilloscope will restart and begin installing the
required software.
NOTE
: If prompted to restart Windows, select Restart Later.
Reinstall the Oscilloscope Application
1. When the X-Stream DSO Setup Wizard appears, click
Next
.
2. On the License Agreement page, select
I Agree
.
3. On the Choose Components page, select the
default (installation)
and click
Install
.
NOTE
: You may see a Windows Security message indicating that Windows can't verify the publisher of
the driver software. Choose "Install this driver software anyway" and click Install.
4. When the X-Stream installation is complete, choose to
Reboot now
and click
Finish
.
Reactivate the F11 Startup Utility
In order for the system recovery wizard to be accessed again from the boot menu, you must reactivate the
F11 startup utility. It is critical to complete these steps after your oscilloscope has restarted.
1. From the Windows
Start Menu
choose
All Programs > Acronis > Acronis True Image Home
.
2. On the Acronis True Image Home window, towards the top right, click
Tools & Utilities
.
3. On the Tools & Utilities page, beneath Protection Tools, click
Acronis Startup Recovery Manager
.
4. On the Acronis Startup Recovery Manager window, click
Activate
. F11 boot time recovery is now ena-
bled.
NOTE
: The process "Searching for Acronis Secure Zone…" may take several minutes.
Restore Software to Current Revision Levels
The recovery process produces a replica of the operating system and oscilloscope application software
at the revision levels that were current when the oscilloscope was manufactured. It does not
automatically upgrade the:
l
Oscilloscope application software (X-Stream)
l
Windows operating system
l
Virus scan definition files
l
Drivers for some hardware options and accessories
Therefore, after completing the disk image recovery, it is highly recommended to search vendor websites
and upgrade the individual components to current revision levels.
The latest oscilloscope application software can be downloaded directly from the Teledyne LeCroy
website at teledynelecroy.com. Most required hardware drivers can be installed during the firmware
upgrade.
Since the calibration data for the oscilloscope is stored in the D: drive, current calibration constants are
not overwritten during recovery of the C: drive. You do not need to restore these.
126
922499 Rev B