Page:
23
A
permeability
test
should
be
performed
after
each
wash
cycle.
The
test
is
performed
as
described
above
for
steps
2
through
6.
Compare
the
flow
rate
obtained
in
the
test
with
previous
benchmark
rates.
If
the
measured
difference
(see
the
calculation
below)
is
15%
or
more
then
further
cleaning
will
be
necessary.
See
the
flow
chart
on
page
24.
To
calculate
the
difference
to
the
benchmark,
do
the
following
calculation:
1.
Subtract
the
measured
flow
rate
from
the
benchmark
flow
rate.
2.
Take
the
result
and
divide
it
by
the
benchmark
flow
rate.
3.
Multiply
the
answer
by
100
and
record
the
result
in
the
Measured
Difference
field
on
the
Permeability
Test
Sheet.