Testing the Switchos
5-2
D850R3
User’s Manual Switchos
5.2
Checking System Components
5.2.1
Solvent Bottle Caps and Degassing Unit
The He lines, the regulating valve and connections can be tested by the following
procedure:
a)
Close the Shut-Off Valves (FIGURE 5-1) and apply 3 bar (45 PSI) of He
pressure.
CAUTION
Caution: Do not operate the He lines a higher pressure than 4 bar (60 PSI).
He Shut-Off
Valve
Vent
FIGURE 5-1
.
The He Shut-Off Valve (Opened Position)
b)
Open the He Regulating Valve on the front panel. The regulating valve should
be turned 90
o
. Check to see if there are any leaks (i.e. are He bubbles
observed in the solvent bottle).
c)
Open the shut-off valve on the bottle cap for solvent bottle A, by closing the
He regulating valve verify that the bubbles of He exit on the side frit of the
sparging/filter frit. Check to make sure that it is possible to regulate the He
flow by opening the He regulating valves until the maximum is reached (90
o
).
d)
Repeat steps (b) and (c) for each solvent bottle.
5.2.2
Solvent Selection Valve – Basic Test
To check the operation of the Solvent Selection Valve and the Valve Controller:
a)
Set the Switchos in local mode by the
Remote/Local
switch on the back
panel (the LEDs of the two push buttons on the front panel should be
illuminated).
b)
Select the next available solvent channel by the
SSV
button. The
corresponding LED should be illuminated and the sound of the moving
solvent selection valve should be heard for a few seconds.
c)
Repeat step (a) for each solvent channel.
Содержание SWITCHOS II
Страница 1: ...D850R3 www lcpackings com Switchos II Advanced Microcolumn Switching Unit User s Manual P N 162013...
Страница 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 10: ...Instructions for Returning Instruments viii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...