User’s Manual Switchos II
D850R3
7-1
Specifications
7.1
Analytical Specifications
7.1.1
10-Port Switching Valves
Type
low dispersion 10-port valve, 2x
Max. Pressure
350 bar (5000 psi)
Wetted Surfaces
standard version
inert version
stator:
stainless steel
PAEK
rotor:
Valcon H
Valcon E
Flow Passage
Diameter
0.15 mm
Switching time
< 100 ms
Operation
local control (push button),
remote control (relay, open collector or TTL)
Position Indicator two LEDs per valve
7.1.2
Solvent Selection Valve
Wetted Surfaces
stator: PPS (polyphenylene sulfide)
rotor: Valcon
E2
Switching Time
< 165 ms (channel to channel)
Position Indicator four LEDs
7.1.3
Loading Pump
Pump Type
reciprocating double piston pump
Flow Rate Range
10 - 500 µL/min
Flow Rate
Accuracy
< 0.5%
Pressure Range
0 - 400 bar
Pressure
Monitoring
indication of µ-column pressure
Rapid Purge Valve yes
Instrument
Control
CHROMELEON
®
software or standalone
CHAPTER 7
Содержание SWITCHOS II
Страница 1: ...D850R3 www lcpackings com Switchos II Advanced Microcolumn Switching Unit User s Manual P N 162013...
Страница 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 10: ...Instructions for Returning Instruments viii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...