Maintenance and Troubleshooting
User’s Manual Switchos II
D850R3
6-13
TABLE 6-3 Typical System Parameters of a Pre-concentration Setup
Typical values (a)
Parameter
CAP Configuration
NAN Configuration
Micropump Flow
(Master flow)
200 µL/min
187.5 µL/min
Micropump Pressure
(Master pressure)
160 – 200 bar
160 – 200 bar
CRP (b)
50
625
Column Flow
4 µL/min
300 nL/min
Column Pressure
60 – 80 bar
100 – 130 bar
Trap Column Flow
30 µL /min
30 µL /min
Trap Column Pressure
40 – 70 bar
40 – 70 bar
Connecting Tubing I.D.
(c)
50 µm
20 µm
Response (delay) Time
‘Injection to injection peak’ (d) 2 -5 min
2 – 5 min
He pressure
1 bar
1 bar
Notes:
a) 100 % mobile phase A (start condition of the UltiMate standard system
checkout procedure with the cytochrome c test sample).
b) No flow sensor for automatic CRP calibration used.
c) Used in the micro flow path.
d) If a trap column is used in conjunction with the Switchos, this is the
time after switching the trap column into the micro flow.
Содержание SWITCHOS II
Страница 1: ...D850R3 www lcpackings com Switchos II Advanced Microcolumn Switching Unit User s Manual P N 162013...
Страница 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 10: ...Instructions for Returning Instruments viii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...