Installation and Getting Started
2-4
D850R3
User’s Manual Switchos
2.3.2
RS-232 Connectors
The two RS-232 serial interfaces enable digital data transfer between the loading
pump, the UltiMate Micropump, the UltiMate UV Detector (if present) and the
PC. These devices communicate with each other to form an integrated network.
To set up the RS-232 network connections:
•
Connect the RS-232
1
connector of the Switchos loading pump, the RS-232
2
connector of the UltiMate UV Detector and the COM port of the PC using
the Y-cable (item a, FIGURE 2-2).
•
Connect the RS-232
2
connector of the Switchos loading pump and the RS-
232
1
connector of the UltiMate Micropump using the Serial Communication
Cable (item b, FIGURE 2-2).
•
Connect the RS-232
2
connector of the Micropump and the RS-232
1
connector of the UV Detector with the same type of cable.
Switchos II
UltiMate
PC COM port X (SW II Valve Control)
PC COM port Y (Solvent Organizer)
PC COM port Z (ULT/SWII Network)
Item
P/N
Cable
a 160069
9 pin
female
9 pin
female
9 pin
male
b 160070
9 pin
male
9 pin
female
1:1
c 160071
9 pin
female
9 pin
female
Null-Modem
a
b
b
c
c
FIGURE 2-2 Setting up the RS-232 Network Connections
If your configuration does not include an UltiMate UV Detector, connect the
y-cable to the RS-232
1
connector of the UltiMate Micropump, the RS-232
2
connector of the Switchos loading pump and the COM port of the PC. Use a
Serial Communication Cable to connect the other RS-232 connectors of both
pumps.
CAUTION
Caution: The RS-232 sockets are to be used only with LC Packings software
products (e.g. CHROMELEON) or third vendor software that support the LC
Packings UltiMate System.
Содержание SWITCHOS II
Страница 1: ...D850R3 www lcpackings com Switchos II Advanced Microcolumn Switching Unit User s Manual P N 162013...
Страница 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 10: ...Instructions for Returning Instruments viii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...