Installation and Getting Started
User’s Manual Switchos
D850R3
2-5
2.3.3
COMMUNICATION Connector
Connect the COMMUNICATION port to a free COM port on the computer using
the Solvent Organizer Com cable (item c, FIGURE 2-2).
2.3.4
INPUTS Connector
CHROMELEON controls the valve positions via the COMMUNICATION port
(Section 2.3.3). Refer to Sections 2.4 and 2.5 if the valves need to be controlled
by digital input signals (e.g. the instrument is to be installed in conjunction with a
software package which does not support the serial communication).
2.3.5
Triggering a Mass Spectrometer (or an Other Device)
If you want to provide a mass spectrometer (or other external device) with a
trigger signal from the CHROMELEON software at some event (e.g. on the
beginning of the gradient program), connect the appropriate input of the mass
spectrometer (or other device) to the EVENT 8 output (relay contact) of the
UltiMate Micropump as shown in FIGURE 2-3.
UltiMate
Micropump
EVENT 8
(Relay)
Mass Spectrometer
or
other device
{
FIGURE 2-3 Cable to trigger the MS (or another Device) by EVENT 8
2.3.6
Power Connector
Since the Switchos is fitted with a universal power supply for input voltages
from 90 to 260 V, manual setting of the supply voltage is not required. The
power cord should be inserted in the socket directly below the Main Power
switch on the right side of the rear panel (FIGURE 2-1). In addition, the loading
pump should be plugged into the socket above the Main Power switch.
DANGER
Danger: The Switchos must be connected to a power source that is connected to
a true ground.
Содержание SWITCHOS II
Страница 1: ...D850R3 www lcpackings com Switchos II Advanced Microcolumn Switching Unit User s Manual P N 162013...
Страница 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 10: ...Instructions for Returning Instruments viii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Страница 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...