4.1.5 MOTOR OPERATION
•
Manually turn chuck and check for free movement and no binding. The motor
should never be overloaded or held from turning.
•
If RPM display shows zeroes when the chuck is actually spinning, this would
indicate failure of the RPM sensor and an E08 error will appear.
•
Deceleration of the motor from a high speed is usually far less controllable than
acceleration. When programming such steps, it is necessary to select a
deceleration rate that the motor can achieve. This usually requires setting the
acceleration index in the 1 to 5 range. If this is not done the momentum of the
chuck and motor will not allow the speed to follow the required motor speed
profile, causing an error. If this occurs, decrease the acceleration index in the
program step that goes to the lower speed. See Section 3.5.1 “PROGRAMMING
ACCELERATION/DECELERATION” of this manual for details.
The fragment adapter must
be installed with a single
o-ring for fragments.
The inside o-ring is for
fragments ≥ 11mm x 11mm
.
The outside o-ring is for
fragments ≥ 20mm x 20mm
.
Only 1 size o-ring should be
installed in the adapter at
any one time.
Improper installation of two
simultaneous o-rings on the
fragment adapter
.