materials could not be combined. Automatic shutoff capability should be
considered for bulk distribution systems. The secondary containment provided
should be able to contain 110% of the volume of the largest single container or
all vessels combined.
2.2 INSTALLATION
SEE FRONT OF MANUAL FOR “QUICK START INSTALLATION INSTRUCTIONS”
.
The first step in the installation is to unpack the spin processor. Identify each
component on the packing list and verify that it was not damaged during shipping
. Any
loss or damage must be reported to the factory within 30 days from shipment.
Refer to section 2.1 “Facility Requirements” for required electrical power, vacuum, drain,
exhaust, and N
2
/CDA (Clean Dry Air). Since each spin processor is configured to meet
the needs of the end-user, requirements will vary for each application. It is
recommended that a point
-
of
-
use filter (≤1 micron), a pressure regulator and a check
valve (5 PSI breaking pressure) be installed on all supply tubing connected to the spin
processor. Refer to installation drawing(s) in Section 5 for guidance.
2.2.1 SPIN PROCESSOR CONNECTIONS
•
Follow the outline in section 2 “Facilities” to prepare for installation. Use
dedicated regulators and filters where appropriate.
•
See “Quick Start Instructions” in front of manual for step-by-step
installation instructions
.
•
Install the processor on a stable, level and vibration free surface.
•
Connect drain line to rear of spin processor’s drain adapter.
See Quick Start
Instructions.
•
Connect vacuum source, using the provided 3/8” (9.52mm) O.D. x ¼”
(6.35mm) I.D. tubing, to the 3/8” fitting on the pneumatic vacuum control valve
located on the right rear side of spin processor. (Applicable only if vacuum-
type chuck is used.)
Note: vacuum supply tubing should be ¼” I.D. or
larger to provide sufficient volume quickly.
See Quick Start
Instructions.
Note:
The 400A model has an external 24V vacuum solenoid.
See
Quick Start Instructions.
•
Connect ¼” O.D. tubing for Seal Purge to N2 / CDA regulator. Set regulator to
60 -70 psi.
See Quick Start Instructions.
•
Connect power cord to an appropriate AC source. Note: the circuit breaker,
located on the back of the processor, is also a power switch, which when
pressed and the button is extended, will turn the system off.