Laurell Technologies
Operations Manual WS-400 Lite Series
All information contained in this manual is the property of Laurell Technologies Corporation® and is NOT to be edited, reproduced or
distributed without express written permission from a corporate officer.
32
3.1 POWER
All units are equipped with a single pole over current circuit breaker with manual reset.
This breaker is located on the body of the spin processor housing near the AC cord
entry location (Indeck systems on the remote controller’s rear panel). Seeing the white
band of the “reset” button is indicative that the power has been interrupted. During
normal operation, the circuit breaker should never trip. If the circuit breaker does trip, a
visual and mechanical inspection of the unit should be performed before resetting the
circuit breaker.
This circuit breaker also serves as an “ON/OFF” switch for the unit. Pressing the “reset”
button will toggle power on and off to the unit.
3.2 INTERLOCKS
3.2.1 LID INTERLOCK
A hidden lid interlock switch disables the spin processor’s motor operation when the lid
is in the open position. If the lid is opened while running a program, the program will be
interrupted and the chuck rotation will stop slowly. This condition is noted by a
flashing
“LID”
indicator on the second line of the LCD display. The program can be continued
from the point at which it stopped by closing the lid and pressing the
RUN/STOP
key.
(See Figure 3-2)
Figure 3-2
FLASHING LID DISPLAY
Flashing LID