![LANDTEC GEM2000 Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/landtec/gem2000/gem2000_operating-manual_3535195079.webp)
GEM2000 Range Gas Analysers
OMGEMN1.41
Copyright 2009 Geotechnical Instruments (UK) Limited
Page 79
12.0 Technical Specifications
12.1 GEM2000 Technical Specification
GEM2000
Technical Specification
POWER SUPPLY
Battery type
Rechargeable Nickel Metal Hydride battery pack containing six 4AH cells
(not user replaceable)
Battery life
Typical use 10 hours from fully charged
Battery lifetime
Up to 1000 charge/discharge cycles
Battery charger
Separate intelligent 2A battery charger powered from mains supply (100-
240V 4763Hz)
Charge time
Approximately 2 hours from complete discharge
Alternative power
Can be powered externally for fixed in place applications. Contact
Geotechnical Instruments (UK) Ltd for further information
Memory backup battery
Lithium Manganese for data retention.
GAS RANGES
Gases measured
CO2 and CH4
By dual wavelength infrared cell with reference channel
O2
By internal electrochemical cell
Other gases can be measured by external gas pods.
Range
CH4
0 -70% to specification, 0-100% reading
CO2
0 -60% to specification, 0-100% reading
O2
0 -25%
Typical accuracy
Gas
0-5% vol
5-15% vol
15%-FS
FS
CH4
±0.5% (vol)
±1.0%
(vol)
±3.0% (vol)
70%
CO2
±0.5% (vol)
±1.0% (vol)
±3.0% (vol)
60%
O2
±1.0% (vol)
±1.0% (vol)
±1.0% (vol)
25%
Response time, T90
CH4
CO2
O2
≤
20 seconds
≤
20 seconds
≤
20 seconds
Oxygen cell lifetime
Approximately 3 years in air
FACILITIES
Temperature measurement
-10°C to +75 ºC with optional probe
Temperature accuracy
± 1.0 ºC with optional probe
Visual and audible alarm
User selectable CO2, CH4 and O2 alarm levels *
Communications
Via USB lead to 7 pin Lemo connector *
Relative pressure
± 500 mbar from calibration pressure
Relative pressure accuracy
± 4mbar typically (should be zeroed before reading) to
± 15mbar max
Available Memory
1000 IDs *
2000 Readings
1000 Events *