![LANDTEC GEM2000 Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/landtec/gem2000/gem2000_operating-manual_3535195077.webp)
GEM2000 Range Gas Analysers
OMGEMN1.41
Copyright 2009 Geotechnical Instruments (UK) Limited
Page 77
10.0 Service
The GEM2000 range analyser should be regularly serviced to ensure correct and accurate
operation. Geotech (UK) Limited recommends a service and recalibration every
6 months
.
The GEM2000 range is ATEX certified for use in potentially explosive areas. As such it should
be serviced only by qualified engineers. Failure to observe this will result in the warranty
becoming invalid and could invalidate the ATEX certification.
Warning
If the GEM2000/GEM2000 Plus analyser is serviced by unqualified
engineers the ATEX certification may be invalidated and the
instrument may be unsafe for use in a potentially explosive
atmosphere.
User serviceable parts:
There are no user serviceable parts inside the instrument.
The following parts can be user serviced:
In-line water filter
This should be regularly inspected for obstructions or damage
and changed if needed. The instrument should never be
operated without the in-line water filter as this may result in
water entering the instrument.
Inlet port filter
The inlet port (particulate) filter is located on the underside of
the instrument and is accessed by removing the plastic plug.
This can be removed by unscrewing using a suitable sized
coin. The filter should be replaced if it is contaminated. Never
operate the instrument without the particulate filter. When
replacing the plug be sure not to over-tighten it - finger tight
is adequate.
Sample tubing
Always ensure that sample tubes are not contaminated or
damaged.
QRC connectors
Periodically check that the O-rings on the QRC gas connectors
are not damaged. A damaged O-ring can let air into the
sample gas and result in incorrect readings. If the O-ring is
damaged the complete QRC connector should be replaced.
H
2
S filter material
When the filter material changes colour to a
light grey
colour
then the filter should be replaced.