GEM2000 Range Gas Analysers
OMGEMN1.41
Copyright 2009 Geotechnical Instruments (UK) Limited
Page 21
6.4 Keypad Lock
Keypad lock avoids accidental key presses when carrying the analyser.
1)
To lock the keypad, press and hold the
‘
Backlight’ key for two seconds.
Note: A message stating that the keypad is locked will be displayed at the bottom of
the analyser read screen.
2)
To unlock the keypad, press and hold the ‘Backlight’ key again.
6.5 Instrument Main Gas Read Screen
The ‘Main Gas Read Screen’ is considered to be the normal operating screen and all
operations are carried out from this starting point.
The actual data shown on this display will depend on the version of the instrument and the
options that have been selected. In general, all of the main readings will be shown.
The instrument will only connect to the Gas Analyser Manager whilst in this screen.
Example - GEM2000 Plus:
6.6 Cold Start
A cold start should only be carried out to correct an instrument if no other course of action
has proved successful, as this function will clear the instrument memory entirely, reset all
factory settings and reset the internal time and date to a default setting.
Note:
This function should only be used as a last resort. For gas calibration error
messages ensure a factory setting and user calibration have been carried out
first.
6.6.1
Perform a Cold Start
Note Please contact Technical Support at Geotech (UK) Limited on +44(0)1926
338111 before performing this operation.
1)
Turn the analyser on and during the self-test continue to hold the ‘Enter’ key until
such time as the self-test has been completed.
Battery
Power
Indicator
Memory
Indicator
Time / Date Stamp
Pump On /Off
Pump Run
Time
Borehole ID
Last Readings
Menu
Options
Available
Battery
Power
Indicator
Memory
Indicator
Time / Date Stamp
Pump On /Off
Pump Run
Time
Borehole ID
Last Readings
Menu
Options
Available