OMGEMN1.41
GEM2000 Range Gas Analysers
Page 66
Copyright 2009 Geotechnical Instruments (UK) Limited
Step 16
Go on to the Calibrate O
2
with Air Process @20.8%
Calibration equipment is not required to perform this operation.
Step 17
Remove gas source if attached.
Step 18
Press
to switch on pump.
Allow the pump to run for approximately two minutes or until reading
stabilises; this purges the unit of gas.
Step 19
Press
to switch off pump.
Step 20
From the ‘Check Calibration’ screen:
Allow the reading to stabilise; this should be approximately 20.8%.
Step 21
Press
to enter ‘Calibration Menu’.
Scroll
to ‘Span Channel(s)’.
Press
to select ‘Span Channel(s)’.
Scroll
to ‘Span O
2
@ 20.8%’ (the value set previously).
Press
to select ‘Span O
2
@ 20.8%’.
‘CALIBRATION COMPLETE’ should be displayed.
Step 22
END