GEM2000 Range Gas Analysers
OMGEMN1.41
Copyright 2009 Geotechnical Instruments (UK) Limited
Page 39
Instructions for Use:
The following diagram shows how to modify the standard water trap and sample tubing
to fit the H
2
S filter.
Note:
When onsite the site engineer must have an unmodified water trap assembly
in addition to the modified water trap with a H
2
S filter in order to take
readings with and without a filter.
1)
Once the H
2
S filter is fitted refer to the process diagrams for ‘Taking Gas and Flow
Measurements’ for details of taking readings using a H
2
S filter.
2)
The analyser can be set to prompt the site engineer when to remove the H
2
S filter
and take a valid H
2
S reading.
Note:
This is only for the GEM2000 Plus.
Sample
Tubing
Attached to
Monitoring
Point or
Sample
Tubing
Cut the
Sample
Tubing
Water
Trap
Insert
H
2
S
Filter
QRC
Connect to
Analyser
Standard Water Trap and Sample Tubing
Modified Water Trap and Sample Tubing with H
2
S Filter
QRC
Connect to
Analyser
Sample
Tubing
Attached to
Monitoring
Point or
Pitot Tube
Sample
Tubing
Cut the
Sample
Tubing
Water
Trap
Insert
H
2
S
Filter
QRC
Connect to
Analyser
Standard Water Trap and Sample Tubing
Modified Water Trap and Sample Tubing with H
2
S Filter
QRC
Connect to
Analyser
Sample
Tubing
Attached to
Monitoring
Point or
Sample
Tubing
Cut the
Sample
Tubing
Water
Trap
Insert
H
2
S
Filter
QRC
Connect to
Analyser
Standard Water Trap and Sample Tubing
Modified Water Trap and Sample Tubing with H
2
S Filter
QRC
Connect to
Analyser
Sample
Tubing
Attached to
Monitoring
Point or
Pitot Tube
Sample
Tubing
Cut the
Sample
Tubing
Water
Trap
Insert
H
2
S
Filter
QRC
Connect to
Analyser
Standard Water Trap and Sample Tubing
Modified Water Trap and Sample Tubing with H
2
S Filter
QRC
Connect to
Analyser