7.0 Taking Readings
7.1 Preliminary Checks Before Taking Readings (Best Practice)
Prior to use, it is good practice to ensure
that:
Step 1
All necessary ID codes and
readings have been uploaded from
GAM to the analyser.
Step 2
Check the ‘Mode of Operation’ is
correct. Choose either GEM2000
for gas extraction monitoring
analyser or GA2000 range for
landfill gas analyser.
Step 3
The instrument has the correct
time and date set.
Step 4
The water trap filter is fitted and is
clean and dry.
Step 5
The inlet port micro-fibre filter is
fitted and is clean and dry.
Step 6
Spare filters are available in case
of accidental water blockage or
contamination.
Step 7
The battery has a good charge
(minimum 25% charge, even if
only a few readings are required).
Step 8
The memory has sufficient space
available.
Step 9
The gases CH
4
, H
2
S and CO
readings have been auto-zeroed,
without gas concentration present.
Step 10
If necessary check the span
calibration with a known
concentration calibration gas.
Step 11
Take readings.
Note: If you are using the optional gas
pod you should zero it before use.