19
• Always immerse the probe beyond upper steel band for best
results. Use the
fi
ll line on the outside of the probe guard for
reference.
4.4 Using the Manual Ranging Function
By default the meter has auto-ranging ability and automatically selects
the range in which the readings appear.
Alternatively, to override the auto-ranging function, a speci
fi
c range can
be manually selected by pressing
successively for each measurement
range. The
fi
ve ranges are:
Conductivity
Range ( CON 6 PLUS)
TDS Range ( TDS 6 PLUS)
(using 0.5 TDS factor)
0 – 20.00
μ
S/cm
0 – 10.00 ppm
0 – 200.0
μ
S/cm
0 – 100.0 ppm
0 – 2000
μ
S/cm
0 – 1000 ppm
0 – 20.00 mS/cm
0 – 10.00 ppt
0 – 200.0 mS/cm
0 – 100 ppt
NOTE:
• If the value of the solution that is being measured is higher than
the range selected, [
Or
] (over range) will appear. Press
to
select a measurable range.
• The meter will reset to the Auto-ranging function when the meter
is turned o
ff
.
S
μ
mS
mS
Auto-ranging
Manual ranging: 0 - 20.00 uS/cm
Manual ranging: 0 - 200.0 uS/cm
Manual ranging: 0 - 2000 uS/cm
Manual ranging: 0 - 20.00 mS/cm
Manual ranging: 0 - 200.0 mS/cm
S
μ
S
μ
S
μ
Содержание CON 6 PLUS
Страница 1: ...1...
Страница 2: ......
Страница 35: ...35...
Страница 36: ...PO Box 329 Chestertown MD 21620 USA 800 344 3100 410 778 3100 f 410 778 6394 www lamotte com 626094 32 11 19...