24
5. Press or
to move to the next setup function or press
CAL
to exit to the measurement mode.
5.6 Normalization Temperature
The meter can be set to normalize the measurements to a temperature of
either 25 °C or 20 °C. The default value is 25 °C.
1. Enter advanced setup as described in Section
5.1.
2. Press or
until [
t.nr °C
] appears. Press
ENTER
.
3. Press or
to select [
25.0 °C
] or [
20.0
°C
].
4. Press
ENTER
to select and return to the [
t.nr
°C
] setup function.
5. Press or
to move to the next setup
function or press
CAL
to exit to the
measurement mode.
5.7 Single-Point Calibration
Select “YES” to apply a single calibration value across
all ranges.
Select “no” to allow separate calibrations for each
range, or to restrict an individual calibration so that it is
applied to one range only. The default value is “Yes”.
1. Enter advanced setup as described in Section
5.1.
2. Press or
until [
S.P.CA
] appears. Press
ENTER
.
3. Press or
to select [
Yes
] or [
no
].
4. Press
ENTER
to select and return to the [
S.P.CA
] setup
function.
5. Press or
to move to the next setup function menu or press
CAL
to exit to the measurement mode.
C
°
C
°
C
°
HOLD
ENTER
HOLD
ENTER
Содержание CON 6 PLUS
Страница 1: ...1...
Страница 2: ......
Страница 35: ...35...
Страница 36: ...PO Box 329 Chestertown MD 21620 USA 800 344 3100 410 778 3100 f 410 778 6394 www lamotte com 626094 32 11 19...