22
HOLD
ENTER
HOLD
ENTER
5.2 Selecting the Cell Constant
The meter includes a probe with a nominal cell constant (k) of 1.0. Use
probes with k = 0.1 and 10 (not provided) for improved performance
in extreme samples. Use this setup function to change the cell constant
if necessary. Meter default is 1.0 to match the probe supplied with the
meter.
k = 0.1 ideal for low measurements <20
μ
S/cm (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS/cm (>10 ppt).
1. Enter advanced setup as described in Section
5.1.
2. Press or
until [
CELC
] appears. Press
ENTER
.
3. Press or
to select “1.0”, “0.1”
or “10.0”. Ensure that the cell constant
corresponds with the probe being used.
4. Press
ENTER
to select and return to [
CELC
]
setup function.
5. Press or
to move to the next setup
function or press
CAL
to exit to measurement
mode.
5.3 Automatic Calibration (CON 6 PLUS only)
Select automatic calibration “YES” to chose four factory pre-set
conductivity calibration standards (see Section
3.3 Table 1
). To use other
standards, or to calibrate any standard manually select “no”. The default
value is “YES”.
1. Enter advanced setup as described in Section
5.1.
2. Press or
until [
ACAL
] appears. Press
ENTER
.
3. Press or
to select [
YES
] or [
no
].
4. Press ENTER to select and return to [
ACAL
]
setup function.
5. Press or
to move to the next setup
function or press
CAL
to exit to the
measurement mode.