3.4 Using Automatic Calibration (CON 6 PLUS) 

In Automatic Calibration mode, the CON 6 PLUS can accept up to 4 

calibration points with a maximum of 1 point per measurement range. 

Note: Values in the 0.00 to 20.00 

μ

S/cm range cannot be calibrated in 

Auto Calibration mode. 

1.  If necessary, press the 

MODE

 key to 

select the conductivity mode.

2.  Rinse the probe with deionized water 

or a rinse solution. Rinse with a small 

amount of calibration standard.

NOTE: For Automatic Calibration one of the 

calibration standards listed in Table 1 must be 

used. 

3.  Dip the probe into the calibration 

standard. Stir gently with the probe to 

create a homogeneous sample. Allow 

time for the reading to stabilize.

4. Press the 

CAL

 key to enter the 

conductivity calibration mode. The 

[

CA

] indicator will appear brie

fl

 y, and 

then a value will appear 

fl

 ashing.

NOTE: To exit calibration without 

con

fi

 rmation, press CAL to return to the 

measurement mode. 

5.  When the value is stable, press 

ENTER

. The calibration standard 

value will appear for 3 seconds. If the 

calibration is successfully performed, 

[

donE

] will be displayed brie

fl

 y before 

meter returns to the measurement 

mode. 

6.  Repeat steps 1-5 as needed with 

additional calibration standards.

 
NOTES: 

•  To protect from erroneous calibrations, the allowable tolerance 

is ±40% of the factory default value. If a calibration is attempted 

with standards that fall outside of the tolerance range, the error 

mS

mS

mS

mS

CAL

HOLD

ENTER

Measurement Mode

Содержание CON 6 PLUS

Страница 1: ...1...

Страница 2: ......

Страница 3: ...without notice as improvements are made and does not represent a commitment on the part of LaMotte Company WARNING Reagents marked with an are considered to be potential health hazards To view or prin...

Страница 4: ...Automatic Temperature Compensation ATC 18 4 2 Without ATC Manual Temperature Compensation 18 4 3 Taking Measurements 18 4 4 Using Manual Ranging Function CON 6 PLUS TDS 6 PLUS 19 4 5 HOLD Function 20...

Страница 5: ...S measures total dissolved solids TDS in parts per million ppm or parts per thousand ppt and temperature C Each meter measures up to 5 di erent ranges and will automatically switch to appropriate meas...

Страница 6: ...abort calibration without confirming value MODE Selects desired measurement mode When pressed simultaneously with ON OFF it will take proceed the SETUP mode See ADVANCED SETUP section for more informa...

Страница 7: ...DS 6 PLUS and C temperature 1 Primary display 2 Parts per million ppm or parts per thousand ppt TDS 6 PLUS 3 Millisiemens cm mS or microsiemens cm S indicator CON 6 PLUS 4 Temperature indicator 5 Perc...

Страница 8: ...e pushing the bottom edge of the meter down to set it into position Lift up the stand at the back of meter for bench top applications Figure 2 2 4 Inserting New Batteries The battery compartment is fo...

Страница 9: ...air entrapment ensuring accurate repeatable and stable readings The wettable materials of the probe include 1 Polyetherimide Ultem protective probe guard 2 Polybutylterphalate Valox sensor housing 3...

Страница 10: ...re information 2 7 Connecting the Probe 1 To connect the probe align the connector slots with the posts of the meter socket and rotate the connector clockwise until it locks 2 To remove the probe rota...

Страница 11: ...2 Advanced Setup to modify 3 The third screen will show the Normalization Temperature which can be set to 25 C or 20 C The default value is 25 C t 25 0 C See Section 5 6 Advanced Setup to modify 4 The...

Страница 12: ...witch between conductivity or TDS measurement mode and temperature measurement mode press the MODE key The annunciator will indicate the measurement mode Measurement Mode C ON OFF ppm pptmV pH MIN MAX...

Страница 13: ...170 0 mS cm TDS 6 PLUS TDS Range Recommended Calibration Solution Range 0 00 10 00 ppm 3 00 to 8 50 ppm 10 0 100 0 ppm 30 0 to 85 0 ppm 100 1000 ppm 300 to 850 ppm 1 00 10 00 ppt 3 00 to 8 50 ppt 10 0...

Страница 14: ...the automatic calibration mode the CON 6 PLUS will automatically select one of four calibration standard values see below depending on the range and normalization temperature being used Automatic cal...

Страница 15: ...Stir gently with the probe to create a homogeneous sample Allow time for the reading to stabilize 4 Press the CAL key to enter the conductivity calibration mode The CA indicator will appear briefly a...

Страница 16: ...ng a 12 00 mS cm conductivity calibration standard 1 If necessary press the MODE key to select the conductivity mode 2 Rinse the probe thoroughly with deionized water or a rinse solution Rinse with a...

Страница 17: ...he temperature measurement mode 3 Press CAL to initiate the temperature calibration CA will appear briefly then a temperature value will start flashing 4 Dip the probe into a solution with a known tem...

Страница 18: ...plied after manually entering the temperature value into the meter The meter will compensate from this fixed value to the normalization temperature Any temperature can be used between 0 and 50 C The d...

Страница 19: ...t range The five ranges are Conductivity Range CON 6 PLUS TDS Range TDS 6 PLUS using 0 5 TDS factor 0 20 00 S cm 0 10 00 ppm 0 200 0 S cm 0 100 0 ppm 0 2000 S cm 0 1000 ppm 0 20 00 mS cm 0 10 00 ppt 0...

Страница 20: ...st 5 ADVANCED SETUP FUNCTIONS 5 1 Advanced Setup Overview Advanced setup allows customization settings such as selecting the electrode cell constant normalization temperature temperature coe cient TDS...

Страница 21: ...alue is Yes CON 6 PLUS only Adjust Temperature Coe cient from 0 0 to 3 0 per C Default value is 2 1 per C C Normalization Temperature Select 20 C or 25 C Default value is 25 C Adjust TDS factor from 0...

Страница 22: ...or to select 1 0 0 1 or 10 0 Ensure that the cell constant corresponds with the probe being used 4 Press ENTER to select and return to CELC setup function 5 Press or to move to the next setup function...

Страница 23: ...factor 4 Press ENTER to select and return to the tdS setup function 5 Press or to move to the next setup function or press CAL to exit to the measurement mode 5 5 Temperature Coefficient The temperatu...

Страница 24: ...to move to the next setup function or press CAL to exit to the measurement mode 5 7 Single Point Calibration Select YES to apply a single calibration value across all ranges Select no to allow separat...

Страница 25: ...tly swirl it while readings are being taken For best results soak a dry probe for at least 5 10 minutes before calibration Rinse the probe with clean water before storing Never scratch the bands with...

Страница 26: ...rate c Make sure sample entirely covers the probe sensors d Move or switch o interfering motor e Replace probe Slow response Dirty Oily probe Clean recondition probe Inaccurate readings can t calibrat...

Страница 27: ...1 to 20 00 0 01 ppt 20 1 up to 200 0 0 1 ppt depending on TDS factor used Accuracy 1 full scale Temperature Resolution Accuracy 10 0 to 110 0 C 0 1 C 0 5 C Cell Constant Selectable 0 1 1 10 0 1 1 10 T...

Страница 28: ...ions in the water increases its conductance considerably the current being carried by the migration of the dissolved ions This is a clear distinction from the conduction of current through metal which...

Страница 29: ...d specific conductivity is Specific Conductivity S C Conductance cell constant k siemens cm cm2 siemens cm where C is the conductance siemens k is the cell constant length area or cm cm2 Since the bas...

Страница 30: ...e relationship between conductivity and temperature is predictable and usually expressed as relative change per degree centigrade This temperature coe cient change per degree depends on the compositio...

Страница 31: ...probes specified for the CON 6 PLUS and TDS 6 PLUS meters should be used These probes have a built in temperature sensor If a probe without a temperature sensor is used the temperature of the solutio...

Страница 32: ...ntly Adjust the temperature coe cient to 0 no compensation by following the instructions in Section 6 5 Temperature Coe cient 2 Wait 5 minutes Stir gently Note T1 and CT1 conductivity at T1 3 Conditio...

Страница 33: ...m for a return authorization number or visit www lamotte com for troubleshooting help The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit...

Страница 34: ...0 US only or 410 778 3100 faxing 410 778 6394 or emailing tech lamotte com Often a problem can be resolved over the phone or by email If a return of the meter is necessary attach a letter with the ret...

Страница 35: ...35...

Страница 36: ...PO Box 329 Chestertown MD 21620 USA 800 344 3100 410 778 3100 f 410 778 6394 www lamotte com 626094 32 11 19...

Отзывы: