COD3 Plus Colorimeter 10.16
Color
COLOR
PLATINUM COBALT METHOD
NO REAGENTS REQUIRED
Color in water may be attributed to humus, peat, plankton, vegetation, and natural
metallic ions, such as iron and manganese, or industrial waste. Color is removed to
make water suitable for domestic and industrial use. Color may have to be removed
from industrial waste before it is discharged to a waterway.
APPLICATION:
Potable water and water with color due to natural
materials.
RANGE:
0–1000 color units
MDL:
20 cu
METHOD:
Color is determined by a meter that has been calibrated
with colored standards of known platinum cobalt
concentration. True color, the color of water in which the
turbidity has been removed, is measured.
SAMPLE HANDLING &
PRESERVATION:
Collect all samples in clean glassware. Determine color
as soon as possible to avoid biological or chemical
changes that could occur in the sample during storage.
INTERFERENCES:
Turbidity will interfere. Filter before testing.
Testing
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...