PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select
Testing Menu
.
3. Select
All Tests
(or another sequence containing
095 Turbidity
) from
Testing
Menu
.
4. Scroll to and select
095 Turbidity
from menu.
5. Rinse a clean tube (0290) with deionized water (turbidity free). Fill to the 10 mL line
with deionized water.
6. Insert tube into chamber, close lid and select
Scan Blank
.
7. Rinse a second clean tube (0290) with sample water. Fill to the 10 mL line with
sample. Cap tube. Wipe off excess water and fi ngerprints. Shake to resuspend
particulate matter. Remove all bubbles before measurement.
8. Insert tube into chamber, close lid and select
Scan Sample
. Record result. Turbidity
measurements should be taken as soon as possible after sample has been
collected.
9. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
NOTE: For the most accurate results, the sample should be at 25 ±4°C.
Turbidity
COD3 Plus Colorimeter 10.16
Testing
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...