COD3 Plus Colorimeter 10.16
Dissolved Oxygen
DISSOLVED OXYGEN
WINKLER COLORIMETRIC METHOD • CODE 3688-SC
QUANTITY
CONTENTS
CODE
30 mL
*Manganese Sulfate Solution
*4167-G
30 mL
*Alkaline Potassium Iodide Azide
*7166-G
30 mL
*Sulfuric Acid 1:1
*6141WT-G
1
Sample Tube, screw cap
29180
1
Cap
28570
*WARNING:
Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Dissolved oxygen is vital to the survival of aquatic organisms. Naturally present,
dissolved oxygen enters the water when plants photosynthesize. Wind and wave action
also cause oxygen from the air to dissolve into water. Dissolved oxygen is consumed
by aquatic animals and by the oxidation, or chemical breakdown, of dead and decaying
plants and animals. The concentration of dissolved oxygen in natural waters can range
from 0 to 14 ppm and is eff ected by temperature and salinity.
APPLICATION:
This method is applicable for the determination of dissolved
oxygen in drinking water, all surface waters and wastewater.
RANGE:
0.0–11.0 Dissolved Oxygen
MDL: 0.6
ppm
METHOD:
This method uses the azide modifi cation of the Winkler
Method with a colorimetric determination of the yellow iodine
produced from the reaction with the dissolved oxygen.
INTERFERENCES:
The presence of other oxidizing agents may cause positive
interferences. Reducing may cause negative interferences.
Nitrite interferences are eliminated with the azide
modifi cation.
Testing
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...