APPLICATION:
Drinking and surface waters, domestic and industrial waste
water.
RANGE:
0.00–3.00 ppm Zinc
MDL:
0.05 ppm
METHOD:
Zinc forms a blue colored complex with Zincon in a solution
buff ered at pH 9.0. Other heavy metals are complexed by
cyanide and the zinc cyanide complex is released by the
addition of formaldehyde before the other metal cyanide
complexes are destroyed. Sodium ascorbate is added to
reduce the interference of manganese.
SAMPLE HANDLING &
PRESERVATION:
Sample should be analyzed within 6 hours after collection.
The addition of hydrochloric acid will help preserve the
metal ion content, however the acid should be neutralized
before analysis.
INTERFERENCES:
The following ions interfere in concentrations greater than
those listed.
Ion
mg/L
Ion
mg/L
Cd(II)
1
Cr(III)
10
Al (III)
5
Ni(II)
20
Mn (II)
5
Co (II)
30
Fe (III)
7
CrO4(II)
50
Fe (II)
9
Testing
Zinc
COD3 Plus Colorimeter 10.16
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...