COD3 Plus Colorimeter 10.16
Dissolved Oxygen
PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select
Testing Menu
.
3. Select
All Tests
(or another sequence containing
043 Dissolved Oxygen
) from
Testing Menu
.
4. Scroll to and select
043 Dissolved Oxygen
from menu.
5. Rinse a clean tube (0290) with untreated sample water. Fill to the 10 mL line with
sample. This tube is the BLANK.
6. Insert tube into chamber, close lid and select
Scan Blank
.
7. Fill a second tube (0290) to the 10 line with the treated “Fixed” sample. This tube is
the SAMPLE.
8. Remove BLANK from colorimeter, insert SAMPLE tube into chamber, close lid and
select
Scan Sample
. Record result.
9. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
Testing
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...