PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select
Testing Menu
.
3. Select
All Tests
(or another sequence containing
090 Sulfi de LR
) from
Testing
Menu
.
4. Scroll to and select
090 Sulfi de LR
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
Scan Blank
.
7. Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 1.0 mL of
*Sulfi de Reagent A (V-4458). Cap and mix.
8. Add 6 drops of Sulfi de Reagent B (V-4459). Cap and mix. Wait 1 minute. Solution
will turn blue if sulfi des are present.
9. Use the 1.0 mL pipet (0354) to add 2.0 mL of Sulfi de Reagent C (4460). Cap and
mix. Color development is immediate and stable.
10. Insert tube into chamber, close lid and select
Scan Sample
. Record result.
11. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
Testing
COD3 Plus Colorimeter 10.16
Sulfi de
, Low Range
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...