COD3 Plus Colorimeter 10.16
Manganese
, High Range
MANGANESE – HIGH RANGE
PERIODATE METHOD • CODE 3669-SC
QUANTITY
CONTENTS
CODE
10 g
Manganese Buff er Reagent
6310-D
15 g
*Manganese Periodate Reagent
*6311-E
1
Spoon, 0.1 g, plastic
0699
1
Spoon, 0.15 g, plastic
0727
*WARNING:
Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the four
digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a
printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Manganese is present in ground water in the divalent state due to the lack of oxygen.
In surface waters, manganese may be in various oxidation states as soluble complexes
or as suspended compounds. Manganese is rarely present in excess of 1 mg/L. It may
impart an objectionable taste or cause staining problems in laundry, but manganese
levels normally encountered in water seldom produce any health hazards. Manganese is
removed from water by various means, including chemical precipitation, pH adjustment,
aeration, superchlorination and the use of ion exchange resins.
APPLICATION:
Drinking and surface waters, domestic and industrial
wastewaters.
RANGE:
0.0–15.0 Manganese
MDL:
0.3 ppm
METHOD:
Periodate oxidizes soluble manganous compounds into
permanganate.
SAMPLE HANDLING &
PRESERVATION:
Manganese may oxidize readily in a neutral water and
precipitate from solution. It may adhere to or be absorbed
by container walls, especially glass. Acidifi ed samples can
be stored in plastic.
INTERFERENCES:
Reducing substances capable of reacting with periodate or
permanganate must be removed or destroyed before the
periodate oxidation is attempted.
Testing
Содержание COD3 Plus
Страница 2: ......
Страница 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Страница 30: ......
Страница 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Страница 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Страница 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Страница 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Страница 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Страница 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...