VitalsBridge Directions for Use
31
rev. A.5
Note:
The ‘Force SpO2
OFF
’ option will not result in the plethysmograph waveform to disappear or ‘flat line’ on the
patient monitor. Instead, the signal may wander in a manner that allows patient monitors to detect that the SpO
2
probe
is off.
Calibration and Zeroing
It may be necessary to calibrate the vital signs to your specific monitor. For instructions on how to complete the
calibrations from the VitalsBridge Connector Software, see below.
SpO
2
Calibration
The VitalsBridge is pre-calibrated with factory settings for a Nellcor standard probe. SpO
2
calibrations are stored on the
VitalsBridge. When the set SpO
2
value significantly deviates from the value shown on the vitals sign monitor, then it is
recommended to perform a new calibration. First, ensure the SpO
2
probe is placed properly in the SpO
2
adapter and the
VitalsBridge is powered on and connected via the VitalsBridge Connector. Next, under the Configuration Tab, Choose
SpO
2
under “Selected Calibration” and press the NEW button.
Note:
Disconnect the
“mock” SpO
2
probe from the adapter box prior to starting a new calibration.