15.
CLEANING AND DISINFECTION
Prima C/ Prima CS COLPOSCOPE
31
6128000-795
Prima C/ Prima CS
Issue 2.2
Printed on November,2020
Following disinfectants are recommended for cleaning and disinfection.
1. Normal household bleach (Sodium hypochlorite 5%)- strength 5000 ppm(10 parts water 1 part bleach).
2. 70% Isopropyl alcohol.
Procedure:
• Take a muslin cloth.
• Moist it to feel wet
• Disinfect/clean the surface gently.
Note – 1. Use mask and Gloves while performing cleaning and disinfection.
2. While cleaning, muslin cloth should not be dripped wet to prevent seepage and rusting to running/bare parts.
3. Alcohol is flammable,its use as a surface disinfectant should be in well-ventilated spaces only.
The rubber caps, sleeves and grips supplied by labomed are recommend for the following program for
autoclaving:
•
Temperature: 134˚C
•
Time: 10 minutes
•
Instrument: Standard Autoclave