INSTRUCTIONS FOR USING THE MICROSCOPE (continued)
Prima C/ Prima CS COLPOSCOPE
c. HOW TO FOCUS THE OBJECT (Fig. 2.15)
•
Bring object into rough focus using arm movements.
•
Rotate fine focusing knob 2 clockwise or anticlockwise to bring object into sharp focus.
•
If the object is outside the range of the fine focusing knob, bring the complete microscope in the focus
range by any of the following means:
a. Roll the mobile stand forward and backward.
b. Adjust the arm towards left- right and up- down.
c. Adjust the microscope angle left and right. For this loosen the head arm knob and rotate the microscope.
A combination of the above steps will give you the true desired focusing median.
•
On focusing the region of interest, change magnification through the magnification changing knob 1.
(Fig. 2.15)
24
6128000-795
Prima C/ Prima CS
Issue 2.2
Printed on November,2020
1
2